3.2. Remote control commands
794 Basic Titrino
116
C40: Initial measured value in pH "5.12", mV (with U and Ipol)
"41", µA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final
measured value.
C41: End volume with SET in ml, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of a SET titration from the conditioning
in ul/min, "3.5".
C44: (Last measured) temperature in °C. Used for the temperature
compensation in pH measurements.
C45: Start volume with DET, MET, SET in ml, "2.800".
C46: Asymmetry pH of CAL, "6.89".
C47: Relative electrode slope of CAL, "0.9950".
3.2.2.72. Info.TitrResults.FixEP.51.Value read
only
etc. up to .59
Info.TitrResults.pK.61.Value read
only
etc. up to .69
Info.TitrResults.TempVar.C70 read
only/read+write
etc. up to .C79
.FixEP: Fix EP with DET, MET. C5X corresponds to X = 1...9.
.pK: With DET, MET. C6X corresponds to X = 1...9.
.TempVar: Temporary variables in TIP corresponding to the assignments in
the submethods.
3.2.2.73. Info.StatisticsVal.ActN read
only
Info.Statistics.1.Mean read
only
Info.Statistics.1.Std read
only
Info.Statistics.1.RelStd read
only
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results
"3"
Data for MN1:
Mean: Mean value (decimal places as in result)
"3.421"
Std: Standard deviation (1 decimal place more than in result)
"0.0231"
RelStd: Relative standard deviation (in %, 2 decimal places)
"0.14"
3.2.2.74. Info.SiloCalc.C24.Name read
only
Info.SiloCalc.C24.Value read
only
Info.SiloCalc.C24.Unit read
only
for
.C25 as for .C24
Info.SiloCalc.C26.ActN read
only
Info.SiloCalc.C26.Mean read
only
Info.SiloCalc.C26.Std read
only
Info.SiloCalc.C26.RelStd read
only
for
.C27 as for .C26
The current values from the silo calculations. C26 is the mean value out of the
C24 variables; C27 comes from C25.
$Q sends:
C24.Name: Name of the assigned value
"RS1"