11 data output – Metrohm 794 Basic Titrino User Manual
Page 62

2.11. Data output
794 Basic Titrino
58
2.11 Data output
Reports for the output at the end of a determination
With
determination is defined.
The entries are specific to the method and are stored in
the method memory.
report:
Definition of report blocks to be printed automatically at
the end of the determination.
The display texts of the Titrino are shown to the left.
The values are the default values.
>report
report:
report:full;curve
Report sequence (input range depends on the mode:
DET: full, short, mplist, curve, derive, comb, scalc full,
scalc srt, param, calc, calib, ff
MET: full, short, mplist, curve, scalc full, scalc srt,
param, calc, calib, ff
SET, MEAS, CAL:
full, short, scalc full, scalc srt, param, calc, calib,
ff
TIP: full, short, scalc full, scalc srt, param, calc, ff
Select a block with
than one report block, set a ";" as a delimiter between
the blocks.
Meaning of the report blocks:
full
Full result report with raw results, calculations and statistics.
short
Short result report with calculations and statistics.
mplist
Measuring point list.
curve
Titration curve (with DET and MET) or volume vs. time (with SET) or
measured value vs. time (with MEAS)
derive
1st derivative of the titration curve (with DET)
comb
Combined titration curve and 1st derivative (with DET)
scalc full Full report of silo calculations.
scalc srt
Short report of silo calculations.
param Parameter
report.
calc
Report with formulas and operands.
calib Calibration
data.
ff
Form feed on printer.
DEF
(
2
def
>report