Metrohm 794 Basic Titrino User Manual
Page 110

3.2. Remote control commands
794 Basic Titrino
106
Mode.Parameter.Statistics.ResTab.Selected original, delete n,
delete
all
Mode.Parameter.Statistics.ResTab.DelN
1...20
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN: Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.24. Mode.Parameter.Evaluation.EPC DET:
0...5...200
MET
pH:
0.1...0.50...9.99
U,
Ipol:
1...30...999
Upol:
0.1...2...99.9
Mode.Parameter.Evaluation.Recognition.Selected
all, greatest,
last, window, OFF
Mode.Parameter.Evaluation.Recognition.Window.1.LowLim
pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.Evaluation.Recognition.Window.1.UpLim
Input range as LowLim
etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH:
0...±20.00, OFF
U;
Ipol:
0...±2000, OFF
etc. up to 9 fix EP's
Upol: 0...±200.0, OFF
Mode.Parameter.Evaluation.pK ON,
OFF
Parameters for DET and MET: Evaluation of the EP's, see page 32.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in µA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U
and Ipol) or µA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or µA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 34.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. µA (for Upol). Fix EP's are
evaluated until the setting OFF is found.