2.7. Parameters, key

794 Basic Titrino

50

2.7.5 Parameters for TIP

In TIP, several commands and methods can be linked to make a titration procedure. The

TIP sequence can be defined with , see page 62.









sequence

Parameters for the TIP sequence see page 63.

statistics

:

Calculation of mean values and standard deviation,

see also page 55.

preselections

:

ON/OFF of various auxiliary functions such as auto-

matic requests after the start and activate pulse.

>statistics

status: OFF




mean n= 2

res.tab: original






delete n= 1

Statistics calculation

Status of statistics calculation (OFF, ON)

If the statistics calculation is switched off, the following

inquiries regarding the statistics do not appear.

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.


>preselections

req.ident.: OFF





req.smpl size: OFF


meas.mode: OFF


meas.input: 1




Preselections for the sequence

Request of sample identifications after start (id1, id1&2,

all, OFF)

After start, sample identifications can be inquired

automatically: Only id1, id1 & id2, all three id's or no

inquiries.

Request of sample size after the start (value, unit, all,

OFF)

Measured quantity (pH, U, Ipol, Upol, T, OFF)

Quantity for measurements with key .

Measuring input (1, 2, diff.)

Inquiry only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 155.

parameters

>sequence