5 parameters for tip – Metrohm 794 Basic Titrino User Manual
Page 54

2.7. Parameters, key
794 Basic Titrino
50
2.7.5 Parameters for TIP
In TIP, several commands and methods can be linked to make a titration procedure. The
TIP sequence can be defined with
sequence
Parameters for the TIP sequence see page 63.
statistics
:
Calculation of mean values and standard deviation,
see also page 55.
preselections
:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.
>preselections
req.ident.: OFF
req.smpl size: OFF
meas.mode: OFF
meas.input: 1
Preselections for the sequence
Request of sample identifications after start (id1, id1&2,
all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after the start (value, unit, all,
OFF)
Measured quantity (pH, U, Ipol, Upol, T, OFF)
Quantity for measurements with key
Measuring input (1, 2, diff.)
Inquiry only with measured quantities pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 155.
parameters
>sequence