beautypg.com

Metrohm 794 Basic Titrino User Manual

Page 57

background image

2.8 Result calculations

794 Basic Titrino

53

RS1 text RS1


RS1 decimal places 2

RS1 unit: %

Text for result output (up to 8 characters)

Text input see page 8.

Number of decimal places for result (0...5)

Selection of result unit (%, ppm, g/L, mg/mL, mol/L,

mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to

6 characters).

Enter next formula, e.g. for RS2.

Meaning of the calculation variables CXX:

C00

Sample size, see page 69.

C01...C19

Method specific operands, see page 54. They are stored with the

method in the method memory.

C21...C23

Sample specific operands, see page 69ff.

C26, 27

Mean values from silo calculations.

C30...C39 Common

variables.

C40

Initial measured value of the sample, last measured value in MEAS.

C41 End

volume.

C42 Determination

time.

C43

Volume drift for SET with conditioning.

C44 Temperature.

C45

Dispensed start volume.

C46 Asymmetry-pH

(calibration).

C47

Electrode slope (calibration).

C51...C59

Fix EP for DET and MET.

C61...C69

pK/HNP values for DET and MET.

C70...C79

Temporary variables for calculations in TIP.