Metrohm 794 Basic Titrino User Manual
Page 108
3.2. Remote control commands
794 Basic Titrino
104
Mode.Parameter.TitrPara.VStep 0...0.10...9.999
.MptDensity: Parameter for DET: Measuring point density.
.MinIncr: Parameter for DET: Minimum increment in µL. If the minimum in-
crement is set to 0, measured values are stored vs. time.
.VStep:
Parameter for MET: Volume increment in mL. With "0", there is no
dispensing and measured values vs. time are entered in the
measuring point list.
3.2.2.13. Mode
.
Parameter.TitrPara.DosRate 0.01...150, max.
Parameters for DET and MET: Dispensing rate for the volume increments in
mL/min. Max. means maximum possible dispensing rate with the Exchange
Unit in current use.
3.2.2.14. Mode.Parameter.TitrPara.SignalDrift pH,U,Ipol:0.5...50...999, OFF
Upol:
0.05...50...99.9, OFF
Mode.Parameter.TitrPara.UnitSigDrift read
only
Mode.Parameter.TitrPara.EquTime 0...26...9999, OFF
Parameters for DET and MET: Criteria for the measured value acquisition.
Measured value drift in mV/min (with pH, U, Ipol) or µA/min (with Upol),
equilibration time in s. OFF means that the corresponding criterion is switched
off. If both criteria are OFF, the measured values are acquired immediately af-
ter dispensing.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 31. After it has been edited once, it
remains in force with the set value.
3.2.2.15. Mode.Parameter.TitrPara.StartV.Type abs.,
rel.,
OFF
Mode.Parameter.TitrPara.StartV.V 0...999.99
Mode.Parameter.TitrPara.StartV.Factor
0...±999 999
Mode.Parameter.TitrPara.StartV.Rate 0.01...150, max.
Parameters for DET, MET, SET: Start volume.
If an absolute start volume (abs.) has been selected, the volume in mL is
valid.
A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size
*
factor
The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.
3.2.2.16. Mode.Parameter.TitrPara.Pause
0...999 999
Parameters for DET, MET, SET: Pause time in s. Is waited off after the dis-
pensing of the start volume.
3.2.2.17. Mode.Parameter.TitrPara.MeasInput
1, 2, diff.
Mode.Parameter.TitrPara.Ipol -127...1...+127
Mode.Parameter.TitrPara.Upol -1270...400...+1270
Mode.Parameter.TitrPara.PolElectrTest ON, OFF
Parameters for DET, MET, SET: