Page 108

background image

3.2. Remote control commands

794 Basic Titrino

104

Mode.Parameter.TitrPara.VStep 0...0.10...9.999

.MptDensity: Parameter for DET: Measuring point density.

.MinIncr: Parameter for DET: Minimum increment in µL. If the minimum in-

crement is set to 0, measured values are stored vs. time.

.VStep:

Parameter for MET: Volume increment in mL. With "0", there is no

dispensing and measured values vs. time are entered in the

measuring point list.


3.2.2.13. Mode

.

Parameter.TitrPara.DosRate 0.01...150, max.

Parameters for DET and MET: Dispensing rate for the volume increments in

mL/min. Max. means maximum possible dispensing rate with the Exchange

Unit in current use.

3.2.2.14. Mode.Parameter.TitrPara.SignalDrift pH,U,Ipol:0.5...50...999, OFF

Upol:

0.05...50...99.9, OFF

Mode.Parameter.TitrPara.UnitSigDrift read

only

Mode.Parameter.TitrPara.EquTime 0...26...9999, OFF

Parameters for DET and MET: Criteria for the measured value acquisition.

Measured value drift in mV/min (with pH, U, Ipol) or µA/min (with Upol),

equilibration time in s. OFF means that the corresponding criterion is switched

off. If both criteria are OFF, the measured values are acquired immediately af-

ter dispensing.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 31. After it has been edited once, it

remains in force with the set value.

3.2.2.15. Mode.Parameter.TitrPara.StartV.Type abs.,

rel.,

OFF

Mode.Parameter.TitrPara.StartV.V 0...999.99

Mode.Parameter.TitrPara.StartV.Factor

0...±999 999

Mode.Parameter.TitrPara.StartV.Rate 0.01...150, max.

Parameters for DET, MET, SET: Start volume.

If an absolute start volume (abs.) has been selected, the volume in mL is

valid.

A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size

*

factor

The factor is valid.

The dispensing rate in mL/min applies to both cases. Max. means maximum

possible dispensing rate with the Exchange Unit in current use.

3.2.2.16. Mode.Parameter.TitrPara.Pause

0...999 999

Parameters for DET, MET, SET: Pause time in s. Is waited off after the dis-

pensing of the start volume.

3.2.2.17. Mode.Parameter.TitrPara.MeasInput

1, 2, diff.

Mode.Parameter.TitrPara.Ipol -127...1...+127

Mode.Parameter.TitrPara.Upol -1270...400...+1270

Mode.Parameter.TitrPara.PolElectrTest ON, OFF

Parameters for DET, MET, SET: