beautypg.com

Metrohm 794 Basic Titrino User Manual

Page 31

background image

2.7 Parameters, key

794 Basic Titrino

27

titr.rate max. ml/min

**titr.





signal drift 50 mV/min

**titr.











equilibr.time 26 s

**titr.








start V: OFF





start V 0.0 ml


factor 0


dos.rate max. ml/min

**titr.












Dosing rate for volume increments (0.01...150 mL/min,

max.)

sets "max.".

The maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Drift criterion for measured value acquisition. (input

range depends on the measured quantity:

pH, U, Ipol: 0.5...999 mV/min, OFF

Upol:

0.05...99.9 µA/min, OFF)

sets "OFF".

This type of measured value acquisition is often called

an equilibrium titration.

"OFF" means that the measured value is acquired after

an equilibration time. This may be useful for slow

titration reactions or when the response of the

electrode assembly is slow.

Waiting time for measured value acquisition. (0...9999 s,

OFF)

sets "OFF".

If no new equilibration time has been entered, the

Titrino calculates an equilibration time appropriate to

the drift, see page 31. The measured value is acquired

as soon as the first criterion (drift or time) has been

met.

Type of start volume (OFF, abs., rel.)

"OFF": start volume switched off

"abs": absolute start volume in mL

"rel.": relative start volume to sample size.

If "abs." is set:

Absolute start volume (0...999.99 mL)

If "rel." is set:

Factor for relative start volume (0...

±

999 999).

Calculated as: start V (in mL) = factor∗sample size

Dosing rate for start volume (0.01...150 mL/min, max.)

sets "max.".

Maximum rate depends on the Exchange Unit:

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min