6.3. User methods

794 Basic Titrino

178

Form.Det Determination of the sample.

- Transfer of the end point as temporary variable

to FormolNo.

'pa

794 Titrino 01102 794.0010

date 2002-01-03 time 09:58 0

SET pH Form.Det

parameters

>SET1

EP at pH 8.50

dynamics 1.5

max.rate 10.0 ml/min

min.rate 25.0 µl/min

stop crit: drift

stop drift 20 µl/min

>SET2

EP at pH OFF

>titration parameters

titr.direction: +

start V: OFF

pause 0 s

meas.input: 1

temperature 25.0 °C

>stop conditions

stop V: abs.

stop V 99.99 ml

filling rate max. ml/min

>statistics

status: OFF

>preselections

conditioning: OFF

req.ident: OFF

req.smpl size: OFF

activate pulse: OFF

------------

'de

794 Titrino 01102 794.0010

date 2002-01-03 time 09:59

SET pH Form.Det

def

>formula

>silo calculations

match id: OFF

>common variables

>report

>mean

MN1=RS1

>temporary variables

C70=EP1

------------