2.17 Storing determination results and silo calculations

794 Basic Titrino

75

Starting from the following silo report:

*

*

*

with "match id: off" the following silo calculation report (scalc full) is obtained:

With "match id: id1" the following silo calculation report (scalc full) is obtained:

The short silo calculation report contains only calculations for the current sample.

The mean values of the silo calculations are available for further result calculations as C26

and C27 and can be used in the Titrino in formulas.

Mean value of C24 ⇒ C26

Mean value of C25 ⇒ C27

Important:
• If work is performed with silo calculations, the method name must be entered in the

silo memory.

• Results will be overwritten in the silo recalculation, as long as the silo line is marked

with "/". If you do not wish such an input, e.g. because you work off an urgent sample

between a series, disconnect the silo.

'si
794 Titrino 01102 794.0010
date 2002-01-03 time 08:54 14
>silo
cycle lines: OFF
save lines: ON
sl method id 1/C21 id 2/C22 id 3/C23 C00 C24 C25
+ 1 11-2 A/12 94-09-12 0.233g 0.142ml/min 98.53%
+ 2 0-15 A/13 94-09-12 0.286g 0.9976 NV
+ 3 0-15 A/13 94-09-12 0.197g 0.9947 NV
+ 4 11-2 A/12 94-09-12 0.288g 0.138ml/min 95.75%
/ 5 11-2 A/15 94-09-12 0.263g 0.145ml/min 100.61%

Assignment

for C24 only

:
method id 1/C21 id 2/C22 id 3/C23 mean +/-s n
11-2 * * * Rate 0.142ml/min 0.0035 3
Content 98.30% 2.438 3
0-15 * * * Titer 0.9962 0.00105 2

All samples

which have been
processed with
the same
method are

combined

:
method id 1/C21 id 2/C22 id 3/C23 mean +/-s n
11-2 A/12 * * Rate 0.140ml/min 0.0028 2
Content 97.14% 1.966 2
0-15 A/13 * * Titer 0.9962 0.00105 2
11-2 A/15 * * Rate 0.145ml/min 0.000 1
Content 100.61% 0.000 1

Sample
processed with
the same
method and

having the same
id1 are

combined

:
method id 1/C21 id 2/C22 id 3/C23 mean +/-s n
11-2 A/15 * * Rate 0.145ml/min 0.000 1
Content 100.61% 0.000 1