3.2 Remote control commands
794 Basic Titrino
107
.pK: pK or HNP evaluation. Possible only in pH and U titrations.
3.2.2.25. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Parameters for DET, MET, SET, MEAS: Automatic inquiry after the start of the
determination. From such an inquiry, the determination continues if the re-
quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.84)
or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.26. Mode.Parameter.Presel.ActPuls ON, OFF
for SET:
first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 163.
3.2.2.27. Mode.Parameter.SET1.EP pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.SET1.UnitEp read
only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.28. Mode.Parameter.SET1.Dyn pH:
0.01...20.00, OFF
U,
Ipol:
1...2000, OFF
Upol:
0.1...200.0, OFF
Mode.Parameter.SET1.UnitDyn read
only
Mode.Parameter.SET1.MaxRate 0.01...10...150, max.
Mode.Parameter.SET1.MinRate 0.01...25.0...9999.9
Parameters for SET: Control parameters, see page 42.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or µA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.
.MinRate: Minimum titration rate in ul/min.
3.2.2.29. Mode.Parameter.SET1.Stop.Type drift, time
Mode.Parameter.SET1.Stop.Drift 1...20...999
Mode.Parameter.SET1.Stop.Time 0...10...999, inf
Mode.Parameter.SET1.Stop.StopT 0...99
999, OFF
Parameters for SET: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in ul/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"
means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".