beautypg.com

Metrohm 794 Basic Titrino User Manual

Page 50

background image

2.7. Parameters, key

794 Basic Titrino

46



temperature 25.0 °C

performed.

Temperature (-170.0...500.0 °C)

Temperature is measured at the start if a T sensor is

connected.

This parameter is used for temperature compensation

in pH measurements.

>statistics

status: OFF




mean n= 2

res.tab: original






delete n= 1

Statistics calculation

Status of statistics calculation (OFF, ON)

If the statistics calculation is switched off, the following

inquiries regarding the statistics do not appear.

Mean value calculation from n single results (2...20)

Result table (original, delete n, delete all)

"original":

The original table is used. Deleted

individual results are again incorporated in

the evaluation.

"delete n": Deletion of single results with the index n.

"delete all": The entire table is deleted.

Delete data from sample number n (1...20)

The deleted result is removed from the statistics

calculation.


>preselections

req.ident: OFF





req.smpl size: OFF


activate pulse: OFF

Preselections for the sequence

Request of sample identifications after start of titration

(id1, id1&2, all, OFF)

After start, sample identifications can be inquired

automatically: Only id1, id1 & id2, all three id's or no

inquiries.

Request of sample size after start of titration (value, unit,

all, OFF)

Pulse output on line "activate" (L6, pin 1) of the remote

socket (ON, OFF)

see page 163.