2.7. Parameters, key

794 Basic Titrino

28

pause 0 s

**titr.




meas.input: 1






I(pol) 1 µA

U(pol) 400 mV


electrode test: OFF





temperature 25.0 °C

Waiting time (0...999999 s)

Waiting time, e.g. for equilibration of the electrode after

the start or reaction time after dosing of start volume.

The pause can be aborted with .

Measuring input for pH and U (1, 2, diff.)

Request for measuring input for pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 155.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 µA)

or the

polarization voltage (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive standby state to a

measurement. "OFF" means that the test is not

performed.

Titration temperature (-170.0...500.0 °C)

The temperature is continuously measured if a T

sensor is connected. This parameter is used for

temperature compensation in pH titrations.


>stop conditions



stop V: abs.

**titr.






stop V 99.99 ml

**titr.


factor 999999

**titr.


stop pH OFF

**titr.









Stop conditions for the titration

If several stop conditions have been set, the criterion

which is met first applies.

Type of stop volume (abs., rel., OFF)

"abs.": absolute stop volume in mL.

"rel.": relative stop volume to sample size.

"OFF": stop volume switched off. Stop volume is not

monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...

±

999999)

Calculated as: Stop V (in mL) = factor ∗sample size

Stop at measured value (input range dependent on

measured variable:

pH: 0...

±

20.00, OFF

U, Ipol: 0...

±

2000 mV, OFF

Upol: 0...

±

200.0 µA, OFF)

sets "OFF".

"OFF" means that the criterion is not monitored.