Metrohm 794 Basic Titrino User Manual

Page 42

background image

2.7. Parameters, key

794 Basic Titrino

38







pause 0 s

**titr.




meas.input: 1






I(pol) 1 µA

U(pol) 400 mV



electrode test: OFF




temperature 25.0 °C
cond.

Exchange Unit

max.

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

Pause (0...999999 s)

Waiting time after start volume, e.g. reaction time after

dosing of a start volume. The waiting time can be

aborted with .

Measuring input (1, 2, diff.)

Inquiry only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 155.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 µA)

or the

polarization potential (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive state to a measurement.

"OFF" means that the test is not performed.

Titration temperature (-170.0...500.0°C).

Temperature is measured at the start of the titration if a

T sensor is connected. The value is used for

temperature compensation in pH titrations.