beautypg.com

Select a zero method, Select a zero method 50 – MTS SWIFT 10 MC Sensor User Manual

Page 50

background image

SWIFT 10 MC Sensors

50

Select a Zero Method

Setting up the Transducer Interface

Select a Zero Method

Before you install a transducer and zero it, you must configure the transducer
interface (TI) for the appropriate operating mode.

Equipment required

You will need:

A laptop computer (at test track) or desktop PC with Window 2000 or XP
operating system.

A USB 2.0 communication cable with type A to type B connectors.

SWIFT Transducer Interface Utilities diskette.

Some experience with DOS commands and text editors.

Modes

There are two separate modes for using the transducer interface. The mode you

choose depends on whether you will use the transducer in a spinning (test track
or high rotation laboratory use) application (AngleMode = 0) or non-spinning

(road simulator) application (AngleMode = 1).

If you are using the same transducer with a road simulator that you used

previously on the test track, or vice versa you must download the proper
calibration file and re-zero the transducer.

In either mode, the zero button on the front panel of the TI is used to zero the

angle and balance the bridges.

What you need to do

To change the angle mode used by the TI you will need to:

1. Copy the original calibration file from the CD or diskette that came with the

transducer to the computer.

Note

A separate calibration file was created at the factory for each transducer.
In the next steps, note the serial number of the transducer identified in
the calibration file and the serial number of the TI box that the file was
downloaded to. This information will be used later.

2. Edit the calibration file to select appropriate angle mode: 0 = spinning, 1 =

non-spinning (fixed).

3. Download the modified calibration file from the computer to the TI box.

4. Repeat the process for all of the transducers.

These steps are described in detail in the following sections.