MTS SWIFT 10 MC Sensor User Manual
Page 106
SWIFT 10 MC Sensors
106
Troubleshooting
Output levels are much
higher or lower than
expected
Gains in the TI calibration
file have been overwritten
or modified.
Use the TI2XFER program to upload the
current TI calibration file (refer to
on page 52). Check that the
gain settings match the original calibration file
sent with the transducer. If they are different,
download the original file to the TI, and make
any mode changes needed for your specific
application.
Changing the original calibration file means
you will lose the original calibration data for
your transducer.
Do not make changes to the original
calibration file on the disk.
Make a copy of the original file and make any
changes to the copy.
The data acquisition scales
are set incorrectly.
Check that the data acquisition scales are set
correctly. The full scale calibration range is
shown on the Calibration Report, with 10 V =
Full Scale (unless you requested custom
calibration output before calibration at MTS).
With the TI turned on and all cables from the
transducer to TI connected, use a handheld
DVM to check the output on individual
channels while they are being loaded to make
sure that the output at the TI box and data
acquisition correlate. (Use the BNC connectors
located at the back panel of the TI.)
The transducer was zeroed
with load applied (or a
different load than the
intended tare weight for
non-spinning applications
only).
Rezero the TI, being careful not to touch or load
the transducer during the zeroing procedure.
A zero offset is present in
the data.
See Zero Offset on the previous pages of this
Troubleshooting Guide.
Troubleshooting Guide (part 5 of 10)
S
YMPTOM
P
OSSIBLE
C
AUSES
S
OLUTION
CAUTION