beautypg.com

MTS SWIFT 10 MC Sensor User Manual

Page 103

background image

SWIFT 10 MC Sensors

Troubleshooting

103

Zero Procedure: The Zero
indicators stay on too long,
or they continue to blink
slowly, even after the wheel
has rotated twice.

The angular output signal is

not reaching the TI. If the

encoder signal is not
reaching the TI, it may

eventually time out and red

fail indicator will light.

Check that all cables are attached and

undamaged. In particular, check the main signal

cable from the transducer to the TI.

Check that the encoder output is present. The

angle output signal from the TI should be a 0–5
V sawtooth output while the transducer is

spinning at a constant velocity.

The EncoderSize parameter
in the calibration file is set

incorrectly.

Check that the EncoderSize parameter is set
properly for the slip ring (EncoderSize=2048).

The red fail indicator will
stay lit if there is no encoder

signal. If the encoder is

disconnected after data
collection has started, the

data collection will time out

and blink slowly.

If this signal is not present, check that the slip
ring/encoder assembly has not been damaged. If

you suspect encoder damage, swap the slip ring

assembly with a known functioning unit to
verify it. If the slip ring/encoder assembly is

damaged, call MTS or replace it with a spare

slip ring assembly.

Some or all transducer
output signals read 0 volts
at the TI output even when
a load is present.

The TI is not turned on.

Check that TI is turned on and the green power

indicator is lit.

The transducer signal cable
is not connected.

Check that the cable is connected between the
TI box and the slip ring (or connector housing

for a road simulator).

Output cables are not
connected.

Check that the output connectors are securely
fastened to the data acquisition device.

Cables or connectors are

damaged.

Check all cables and connectors, particularly the

power and transducer cable. Cables must be free
of nicks or cuts, and all connector pins shown

on the cable drawings must be present and not
bent.

Data acquisition

configuration, cabling, or
input ports are incorrect or

damaged.

Verify that the data acquisition is configured

correctly. While the TI is turned on and the
cable from the transducer to the TI is connected,

use a handheld DVM to verify output on

individual channels while they are being loaded.

The wrong calibration file is

loaded in the TI box

Verify the correct calibration file and if

necessary, download the correct file to the TI

box.

Grounding problem.

Check that there is a good signal-to-ground

connection between the TI power cable and the

negative terminal on the battery

Troubleshooting Guide (part 2 of 10)

S

YMPTOM

P

OSSIBLE

C

AUSES

S

OLUTION