3.5.4.2 Sensors Tab
Figure 3-19 Sensors tab
Name: A meaningful name assigned to each sensor. For clear display, keep the 
name to less than 8 characters.
Physical Sensor: Up to six SQM-242 cards (physical sensors 1 to 24) may be 
installed in a computer. However, the SQS-242 software can display a maximum of 
8 sensors simultaneously. Use this dropdown to assign a “physical” sensor to a 
“display” sensor. 
Dual: Indicates that a pair of sensors is set up as primary/secondary duals. When 
a primary sensor fails, the SQS-242 switches to the secondary sensor.
Monitor Output: Select the output that each sensor is positioned to measure. The 
rate and thickness displayed by the sensor will be calculated based on the material 
assigned to the selected output. 
Control Checkboxes: Click Rate to assign the sensor to the PID rate control loop 
for the assigned output during deposition. Click Thk to use the sensor for Thickness 
endpoint detection. Typically you will check both boxes so that the sensor controls 
to rate setpoint and detects the thickness endpoint. If multiple sensors are 
assigned to control the same output, the sensor readings are averaged when 
calculating rate and thickness.
Uncheck both boxes to have a sensor monitor an output, without controlling 
deposition rate or stopping when final thickness is reached.