Temporal Average . . . . . . . . . . . . . Number of measurement samples to
average as the effective rate. More samples
yield a more accurate rate, however, rate and
thickness will not update as quickly at higher
sampling rates.
The effective frequency accuracy is 0.03 Hz
per measurement divided by the square root
of the number of temporal average samples.
Good Rate Tolerance(%) . . . . . . . . Set as a percentage of rate deviation. When
set for good rate, a relay will activate only
within tolerance, when the good rate
condition is true.
For example, if Good Rate Tolerance is set to
10% and the target rate is set to 10 Å/s, when
the rate is between 9 and 11 Å/s (10%), the
good rate relay would be active.
NOTE: A relay must be programmed to use
this feature.
3.2.1.2 System Settings
Operational Mode. . . . . . . . . . . . . . Sets STM-2XM mode of operation (see
section 3.3, STM-2XM Operating Modes, on
page 3-17
Table 3-1 Modes of operation
Mode
Description
Notes
Simple X1
Single film, single sensor
Sensor 1 active only. Analog/shutter outputs
selected by changing active source.
Simple X2
Single film, single sensor
Sensor 2 active only. Analog/shutter outputs
selected by changing active source.
Backup
Single film, dual sensors
Crystal switching mode.
Averaging
Single film, two sensors spatially
averaged
Used for large chambers or systems with a
moving substrate.
Alloy
Two films codeposited
Codeposition of two materials onto a
substrate at the same time.
Independent
Two films cluster deposition
Allows STM-2XM to be used as two separate
instruments. Recommended for cluster tools
and research systems.