9 command: o – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual
Page 111

5 - 31
PN
07
4-
61
4-
P1
A
STM-2XM Operating Manual
5.8.9 Command: O
Command Type . . . . . . . . . . . . . . . . Remote
Command Structure . . . . . . . . . . . . . CEVNTID_class Eid EVENT GROUP
command
Syntax. . . . . . . . . . . . . . . . . . . . . . . . On
Description . . . . . . . . . . . . . . . . . . . . Zeroes thickness, opens or closes a shutter,
selects a sensor if in backup mode, and
forces a final thickness. This command
corresponds with the remote actions using
the front panel buttons, where n is 1 to 13.
Parameters. . . . . . . . . . . . . . . . . . . . (See
.)
Table 5-5 O command
#
Command Structure
Description
1
CEVNTID_AckMemLoss
Ack memory loss state if present
2
CEVNTID_RCzeroch1
Zero Channel 1 thickness
3
CEVNTID_RCzeroch2
Zero Channel 2 thickness
4
CEVNTID_RCGoOpen1
Open Channel 1 shutter function
5
CEVNTID_RCGoOpen2
Open Channel 2 shutter function
6
CEVNTID_RCGoClosed1
Close Channel 1 shutter function
7
CEVNTID_RCGoClosed2
Close Channel 2 shutter function
8
CEVNTID_RCSelectXtl1
Select Sensor 1 in backup mode
9
CEVNTID_RCSelectXtl2
Select Sensor 2 in backup mode
10 CENVTID_RCFnlThkTrgr1
Force final thickness event Channel 1
11 CENVTID_RCFnlThkTrgr2
Force final thickness event Channel 2
12 CENVTID_RCFnlThkTrgrDual
Force final thickness in both channels
13 CEVNTID_RCGoAlloy
Allow sync, channel non-specific
(Alloy mode)
14 CEVNTID_UIFEvnt
User interface event
Reset User Display
Dim down timeout
Illuminate display as though a button was
pressed or the spin knob was rotated