5.8.7 Command: K/L
Command Type . . . . . . . . . . . . . . . . Query
Syntax . . . . . . . . . . . . . . . . . . . . . . . Kn
Description . . . . . . . . . . . . . . . . . . . . Queries the state of relays in Remote mode,
where n is Relay ID 1 to 8.
Command Type . . . . . . . . . . . . . . . . Update
Syntax . . . . . . . . . . . . . . . . . . . . . . . Ln=S
Description . . . . . . . . . . . . . . . . . . . . Updates the state of relays that are in
Remote mode, where n is Relay ID 1 to 8 and
S is state 0 or 1 (False or True, respectively).
NOTE: If a relay is not set to Remote mode, the value set or returned has no effect
on the hardware. The last value sent will be used and the hardware state
of the relay associated with it will change if the relay becomes Remote after
power up.
5.8.8 Command: M/N
Command Type . . . . . . . . . . . . . . . . Query
Syntax . . . . . . . . . . . . . . . . . . . . . . . Mn
Description . . . . . . . . . . . . . . . . . . . . Queries the voltage of analog outputs set to
Remote mode, where n is analog channel 1
to 4.
Command Type . . . . . . . . . . . . . . . . Update
Syntax . . . . . . . . . . . . . . . . . . . . . . . Nn=value
Description . . . . . . . . . . . . . . . . . . . . Updates the voltage of analog outputs set to
Remote mode, where n is analog channel 1
to 4 and value ranges from -10000 to
10000 mV.
NOTE: If an output is not set to Remote mode, the value set or returned has no
effect on the hardware. The last value sent will be used and the hardware
state of the analog output associated with it will change if the output
becomes Remote after power up. The range of values can be expressed
to mV resolution, but STM-2XM will output analog values to the nearest 10
mV.