or retrieve readings in the memory. In addition, the number for STORE
function only can be set through front panel.
INITiate (This command tells the meter to be on “wait-for-trigger” state.
After a measurement is taken, measurement readings will be placed in
the memory.)
FETCh? (Use this command to retrieve stored readings.)
DATA:POINts? (Use this command to query the number of stored
readings.)
4.4.4 Sensitivity Band (Hold)
The reading hold function captures and holds a stable reading on the
display. The multimeter beeps and holds the value when it detects a
stable reading. The sensitivity band in reading hold decides which
reading is stable enough. This band is expressed as a percent of reading
on the selected range. The multimeter considers a reading stable when
three consecutive values are within the band. User has an option to
adjust this sensitivity band.
Default
The default band is 0.1%. User’s selection is stored in a volatile memory
and it will be cleared after the meter has been turned off.
How to adjust the sensitivity band
You can adjust the sensitivity band through either the front panel
operation or the remote interface operation.
Front Panel Operation
In order to adjust the sensitivity band, please press MENU button then
use ◁ and ▷ buttons to find “TRIG” submenu. Press ENTER to select it.
Then again use ◁ and ▷ buttons to find “READ HOLD” submenu. Press
ENTER to choose it. Then use ◁ and ▷ buttons to choose a desired
band. Press ENTER to set the desired sensitivity band. The locations of
the buttons are shown with red rectangle frames in Figure 4-30.