Table 4-4
How to specify a delay time
User can set the delay time from either the front panel operations or the
remote interface operations.
Front Panel Operation
To set the delay time, press MENU button and use ◁ and ▷ buttons to
Measurement Function
Setting
Trigger Delay
Time
PLC >= 1
1.5 ms
DCV/DCI
PLC < 1
1.0 ms
100Ω ~ 100kΩ
1.5 ms
1 MΩ
15 ms
Ω2 and Ω4
(PLC >= 1)
10 MΩ ~ 100 MΩ
100 ms
100Ω ~ 100kΩ
1.0 ms
1 MΩ
10 ms
Ω2 and Ω4
(PLC < 1)
10 MΩ ~ 100 MΩ
100 ms
3 Hz
7.0 s
20 Hz
1.0 s
ACV/ACI
(Remote Interface/
External Trigger/
Single Trigger)
200 Hz
600 ms
3 Hz
1.5 s
20 Hz
200 ms
ACV/ACI
(Front Panel w/ Auto
Trigger On)
200 Hz
100 ms
Remote Interface /
External
1.0 s
Frequency/Period
Front Panel w/ Auto
Trigger ON
0 s