BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 155
156
Set a trigger delay time in seconds. The delay is the time between the trigger
signal and each sample that follows. Specify a delay time from 0 to 3600
seconds.
TRIGger:DELay?
Query the trigger delay time.
TRIGger:DELay:AUTO {OFF|ON}
Disable or enable a automatic trigger delay. The delay is determined by function,
range, integration time, and ac filter setting. Specifying a delay time
automatically turns off the automatic trigger delay.
TRIGger:DELay:AUTO?
Query the automatic trigger delay mode. Returns “0” (OFF) or “1” (ON).
SAMPle:COUNt {<value>|MINimum|MAXimum}
Set the number of readings (samples) the multimeter takes per trigger. Select a
number from 1 to 50,000 readings per trigger.
SAMPle:COUNt ? [MINimum|MAXimum]
Query the sample count.
TRIGger:COUNt {<value>|MINimum|MAXimum|INFinite}
Specify the number of triggers the multimeter will accept before returning to the
“idle” state. Select from 1 to 50,000 triggers. The INFinite parameter instructs
the multimeter to take triggers continuously. Trigger count is ignored while in
local operation.
TRIGger:COUNt? [MINimum|MAXimum|INFinite]
Query the trigger count. Returns “9.90000000E+37” when you select the
infinite trigger count.
C.8 The System-Related Commands
FETCh?
Transfer readings stored in memory by the INITiate command to output buffer