Frequency and period characteristics, Ac characteristics – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 133

134
Frequency and Period Characteristics
Accuracy ± (% of reading)
5
Function
Range
6
Frequency (Hz)
1 Year
23º C±5º C
3-5
0.10
5-10
0.05
10-40
0.03
Frequency
&
Period
100mV
to
750V
40-300K
0.01
AC Characteristics
Accuracy ± (% of reading + % of range)
7
Function
Range
8
Resolution
Frequency
(Hz)
1 Year
(23°C ± 5°C)
3-5
1.00 + 0.04
ACV (AC
TRMS
Voltage)
9
100.0000mV
0.1 uV
5-10
0.35 + 0.04
5 Specifications are for 6 ½ digits and two hours warm up.
6
20% over range on all ranges except 750 Vac range.
7
Specifications are for 6 ½ digits and two hours warm up, slow AC
filter (3 Hz Bandwidth), sine wave input.
8 20% over range on all ranges except 750 Vac range.
9 Specifications are for sine wave input>5% of range. When the inputs
are from 1% to 5% of range and<50 KHz, add the additional error for