beautypg.com

9 remote interface selection, 10 input terminal switch – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 68

background image

69

4.1.9 Remote Interface Selection

The multimeter supports both GPIB and USB interfaces, but only one

interface can be activated at a time. If you are using GPIB, you must set

the address for the multimeter. You can set the address to any value

from 0 and 31. The address is set to “11” when the multimeter is shipped

from the factory.

※ Note: The remote interface can only be set through the front panel

operations.

How to select a remote interface

Press MENU and then use ◁ and ▷ to select INTERFACE. Then Press

ENTER on your choice of USB or GPIB interface.

How to set address when using GPIB

Press MENU and then use ◁ and ▷ to access INTERFACE. Select GPIB

to set address. Use ◁ , ▷ , △ and ▽ to adjust the numbers to the

desired address. Press ENTER to set it.

4.1.10 Input Terminal Switch

Definition

User can select either the front input terminals or the rear input

terminals for their measurements. There are 5 input terminals in the

front panel as well as the rear panel. If the rear terminal is enabled the

REAR annunciator at the upper side will be shown.

Default

The default situation is the front terminals.

How to switch the terminals:

To switch between the front and the rear input terminals, press