3 reading memory (store&recall) – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 94

95
The following commands show how to use the remote interface operation
to disable or enable the beeper:
SYSTem:BEEPer
SYSTem:BEEPer:STATe {OFF|ON}
4.4.3 Reading Memory (Store & Recall)
1201 has a memory capacity of 2000 readings. The readings are stored
in first-in-first-out order and the memory type is volatile, which means
the stored readings will be cleared when the multimeter is power-off. The
reading memory feature can be used for all measurement functions,
math operations, trigger operations and reading hold.
How to use the reading memory
User can store the readings and access to the stored readings through
either the front panel operation or the remote interface operation.
※ Note: Each datum stored from 1201 to remote interface will be in a
first in and first out condition.
Front Panel Operation
Before using the reading memory feature, user needs to select a
measurement function (or the math function) first and then select the
trigger mode.
To store readings, use the following steps:
Press STORE button, and the multimeter will start to store the readings
produced until the specified number of readings is reached. The default
number of readings to store is 100. To change this number, press
CONFIG and then STORE buttons. Then use ◁ and ▷ buttons to move
between the digits and ▽ and △ buttons to increase or decrease
numbers as desired. When ready, press ENTER button. The default
number of readings will be restored when the meter has been turned off.
The location of the buttons are shown with red rectangle frames in
Figure 4-28.