BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 128

background image

129

-148 Character not allowed

A discrete parameter was received but a character string or a numeric

parameter was expected.

-151 Invalid string data

An invalid character string was received.

-158 String data not allowed

A character string was received but not allowed for the command.

-160~-168 Block data errors

Block data is not acceptable.

-170~-178 Expression errors

The meter does not accept mathematical expression.

-211 Trigger ignored

A Group Execute Trigger (GET) or *TRG was received but the trigger was

ignored.

-213 Trigger deadlock

A trigger deadlock occurs when the trigger source is BUS and a READ?

Command is received

-214 Init Ignored

An INITiate command was received but could not be executed because a

measurement was already in progress. Send a device clear to halt a

measurement in progress and place the meter in the “idle” state.

-221 Settings conflict

This error can be generated in one of the following situations:

Situation 1: You sent a CONFigure or MEASure command with

autorange enabled and with a fixed resolution.

Situation 2: You turned math on and then changed to a math operation

that was not valid with the present measurement function.