BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 122

123
accept the trigger. The multimeter takes immediate internal trigger, a
software trigger from a remote interface and external trigger from the
rear panel. Then be sure that the multimeter is ready for a trigger (in the
“wait-for-trigger” state).
The multimeter accepts a trigger only when it is in the “wait-for-trigger”
state. When you have finished configuring the multimeter and have
selected a trigger source, you need to place the multimeter in the
“wait-for-trigger” state so it will take the trigger and make the
measurement. The INITiate, READ? and MEASure? commands all place
the multimeter to the “wait-for-trigger” state.
Triggering Commands
INITiate
READ?
TRIGger:
SOURce {BUS|IMMediate|EXTernal}
SOURce?
TRIGger:
DELay {
DELay? [MINimum|MAXimum]
TRIGger:
DELay:AUTO {OFF|ON}
DELay:AUTO?
SAMPle:
COUNt {
COUNt? [MINmum|MAXimum ]
TRIGger:
COUNt {
COUNt? [MINmum|MAXimum]