BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 100
101
(Slow 5.5 & Fast 6.5) 1
105
19.0
(Slow 6.5) 10
408
4.9
NPLC
Take Time with 2000 Readings(sec)
rate(ch/s)
(Fast 4.5) 0.02
306
6.5
(Slow 4.5 & Fast 5.5) 0.1
318
6.3
(Slow 5.5 & Fast 6.5) 1
442
4.5
Mix function
(VDC+2WRES)
(Slow 6.5) 10
1710
1.2
How to program each channel with measurement function
Press CONFIG + SHIFT + DIGITS for scanning configuration. Use ◁ and
▷ to scroll through submenus. Press ENTER on “SET SCAN CHA”
submenu. Use ◁ and ▷ to switch from channel 1 to channel 10, and
use △and ▽ to switch among the measurement functions. Leave an
unused channel with a blank space (----). Press ENTER when ten
channels are all defined.
How to configure scan count, scan interval and store
measurement readings
Press CONFIG + SHIFT + DIGITS for scanning configuration. Use ◁ and
▷ buttons to scroll through submenus. Press ENTER to select
“OPERATION” submenu. Use ◁ and ▷ to switch among scan count
(COUNT), scan interval (TIMER) and store (STORE) options. Then press
ENTER on your selection. Use ◁ and ▷ to move through the digits and
△and ▽ to increase or decrease the numbers to a desired value. For
STORE option, you can select ON to instruct the meter to store the
measurement results for later retrieving or select OFF .Press ENTER to
set the value.
How to open and close the channels
Press CONFIG + SHIFT + DIGITS for scanning configuration. Use ◁ and
▷ buttons to scroll through submenus. Press ENTER on “----” when you
need to close a specific channel. Press ENTER again on “CHANNEL”. Use
◁ and ▷ to move through the digits and △ and ▽ to increase or
decrease the numbers to the desired channel number. Press ENTER to
close the channel. To open all channels, simply press ENTER on “OPEN
ALL” option. Press SHIFT + DIGITS to scan.