beautypg.com

BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 27

background image

28

DIGITS: Changes resolution.

RATIO: Enables the dcv:dcv ratio function.

%: Calculates the ratio to a target value in percentage.

MIN/MAX: Captures the minimum or maximum readings from the

measurement.

NULL: Activates the offset function in order to get the real measured

reading.

2-4. Second row with SHIFT button:

STEP: Scans from a channel to the next channel in delayed action

when using the scanner card.

SCAN: Enables scanning function when using the scanner card.

LIMITS: Used for Setting upper and lower limit values for readings.

MX+B: Used for calculating slope. X is the normal display reading. M

and B are constants specified by user for scale factor and offset.

dBm: Used for displaying voltage measurement in dBm power unit.

dB: Used for displaying voltage measurement in decibel unit.

2-5. Third row without SHIFT button:

SINGLE: Manually triggers the multimeter to make measurements.

AUTO TRIGGER: Instructs the multimeter to make measurements

continuously.

STORE: Stores a specified number of subsequent readings.

RECALL: Displays stored readings and buffer statistics. Use ◁▷ or

△▽ searching buttons to toggle between reading number and

reading.

LOCAL: Cancels USB or GPIB remote mode.

SHIFT (in blue): Used for accompanying other button with upper print

in blue and converting function.

2-6. Third row with SHIFT button:

EXTRIG: Selects external triggers as the trigger source via BNC port

on the rear panel.

HOLD: Holds reading.

3-1. The first row in SETUP section

◁▷: Scrolls through buffer, conceals or reveals the digits while