beautypg.com

6 stepping (step) – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 101

background image

102

The locations of the buttons are shown with red rectangle frames in

Figure 4-31.

Figure 4-31

4.4.6 Stepping (Step)

Stepping is scanning with a specified time delay between taking input

signal through the defined channels.

※ Note: Setting range of scanner card is only limited on DCV, DCI, ACV,

ACI and Resistance functions.

How to program each channel with measurement function

Press CONFIG + SHIFT + FILTER for scanning configuration. Use ◁ and

▷ to scroll through submenus. Press ENTER on “SET SCAN CHA”

submenu. Use ◁ and ▷ to switch from channel 1 to channel 10, and

use △and ▽ to switch among the measurement functions. Leave an

unused channel with a blank space (----). Press ENTER when ten

channels are all defined.

How to configure scan count, scan interval and store

measurement readings

Press CONFIG + SHIFT + FILTER for scanning configuration. Use ◁ and

▷ buttons to scroll through submenus. Press ENTER to select

“OPERATION” submenu. Use ◁ and ▷ to switch among scan count

(COUNT), scan interval (TIMER) and store (STORE) options. Then press

ENTER on your selection. Use ◁ and ▷ to move through the digits and

△and ▽ to increase or decrease the numbers to a desired value. For