BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 159
![background image](/manuals/545997/159/background.png)
160
*OPC
Sets the “operation complete” bit (bit 0) in the Standard Event register after the
command is executed.
*OPC?
Returns “1” to the output buffer after the command is executed.
*PSC {0|1}
Power-on status clear. Clear the Status Byte and Standard Event
register enable masks when power is turned on (*PSC 1). When *PSC 0
is in effect, the Status Byte and Standard Event register enable masks
are not cleared when power is turned on. [Stored in non-volatile memory]
*PSC?
Query the power-on status clear setting. Returns “0” (*PSC 0) or “1” (*PSC 1).
*SRE
Enable bits in the Status Byte enable register.
*SRE?
Query the Status Byte enable register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
*STB?
Query the Status Byte summary register. The *STB? command is
similar to a serial poll but it is processed like any other instrument
command. The *STB? command returns the same result as a serial
poll but the “request service” bit (bit 6) is not cleared if a serial poll
has occurred.
C.10 SCPI Compliance Information
This section encloses a list of commands that are device-specific to the
1201. Although not included in the 1999.0 version of the SCPI standard,
these commands are designed with the SCPI format in mind and they
follow all of the syntax rules of the standard.