BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 56
57
the stack. In the repeating average mode, the multimeter waits for the
measurement reading stack to fill up and then take an average to
produce a reading for display. It then flushes the stack and starts over
with an empty stack. Consequently, the repeating digital filter yields one
reading for display every specified number of measurement readings.
Digital filter is not available for diode, continuity, frequency and period
measurements.
Default
The digital filter is enabled and is in moving average mode with 10
readings by default.
How to enable/disable digital filter
Press FILTER button to switch the digital filter function. The “FILT”
anunnciator indicates the state of the digital filter. When it is lit, the filter
is enabled.
How to configure digital filter:
You can configure the digital Filter either through the front panel
operation or through the remote interface operation.
Front Panel Operation
For READINGS setting, please press CONFIG and then FILTER button for
selecting READINGS. Press ENTER to access. Use ◁, ▷, △ and ▽
buttons to set the READINGS (from 2 to 100) in order to get an average
value. Press ENTER when it’s ready.
For MODE setting, please press CONFIG and then FILTER button for
selecting MODE. Press ENTER to access. Use ◁, ▷ to select MOVING AVG.
or REPEAT AVG.. Please ENTER to confirm.
Press FILTER button to enable or disable this function.
Remote Interface Operation
Use the following commands to set your digital filter:
SENSe:AVERage:TCONtrol {MOVing|REPeat}
SENSe:AVERage:TCONtrol?