BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 56

background image

57

the stack. In the repeating average mode, the multimeter waits for the

measurement reading stack to fill up and then take an average to

produce a reading for display. It then flushes the stack and starts over

with an empty stack. Consequently, the repeating digital filter yields one

reading for display every specified number of measurement readings.

Digital filter is not available for diode, continuity, frequency and period

measurements.

Default

The digital filter is enabled and is in moving average mode with 10

readings by default.

How to enable/disable digital filter

Press FILTER button to switch the digital filter function. The “FILT”

anunnciator indicates the state of the digital filter. When it is lit, the filter

is enabled.

How to configure digital filter:

You can configure the digital Filter either through the front panel

operation or through the remote interface operation.

Front Panel Operation

For READINGS setting, please press CONFIG and then FILTER button for

selecting READINGS. Press ENTER to access. Use ◁, ▷, △ and ▽

buttons to set the READINGS (from 2 to 100) in order to get an average

value. Press ENTER when it’s ready.

For MODE setting, please press CONFIG and then FILTER button for

selecting MODE. Press ENTER to access. Use ◁, ▷ to select MOVING AVG.

or REPEAT AVG.. Please ENTER to confirm.

Press FILTER button to enable or disable this function.

Remote Interface Operation

Use the following commands to set your digital filter:

SENSe:AVERage:TCONtrol {MOVing|REPeat}

SENSe:AVERage:TCONtrol?