BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 109
110
input. The 500nA ohms current source is connected to produce a nominal
5V signal. A 20ms ADC measurement is performed and the result is
checked against a limit of 5V
1V.
613 Ohms 5 uA source failed This test configures to the 1000V dc
range with the internal 10M 100:1 divider R204 connected across the
input. The 5
A ohms current source is connected. The compliance limit
of the current source is measured. A 20ms ADC measurement is
performed and the result is checked against a limit of 0.12V
0.01V.
614 DC 1000V zero failed This test configures to the 1000V dc range
with no input applied. A 20ms ADC measurement is performed and the
result is checked against a limit of 0V
5 mV.
615 Ohms 10 uA source failed This test configures to the 1000V dc
range with the internal 10M 100:1 divider R204 connected across the
input. The 10
A ohms current source is connected. The compliance limit
of the current source is measured. A 20ms ADC measurement is
performed and the result is checked against a limit of 0.12V
0.01V.
616 DC current sense failed This test configures to the 3A dc range. A
20ms ADC measurement is performed and the result is checked against
a limit of 0A
5A. This test confirms that the dc current sense path is
functional. The test limit is set wide because K303
does not open the
current input during self-test. This test should catch a dc current sense
failure without causing false failures when current inputs are applied
during self-test.
617 Ohms 100 uA source failed This test configures to the 1000V dc
range with the internal 10M 100:1 divider R204 connected across the
input. The 100
A ohms current source is connected. The compliance
limit of the current source is measured. A 20ms ADC measurement is
performed and the result is checked against a limit of 0.12V
0.01V.
618 DC high voltage attenuator failed This test configures to the
1000 V dc range. The 500nA ohms current source is connected to
produce a nominal 5V signal. A 20ms ADC measurement is performed
and the result is checked against a limit of 5V
1V.