input. The 500nA ohms current source is connected to produce a nominal

5V signal. A 20ms ADC measurement is performed and the result is

checked against a limit of 5V

1V.

613 Ohms 5 uA source failed This test configures to the 1000V dc

range with the internal 10M 100:1 divider R204 connected across the

input. The 5

A ohms current source is connected. The compliance limit

of the current source is measured. A 20ms ADC measurement is

performed and the result is checked against a limit of 0.12V

0.01V.

614 DC 1000V zero failed This test configures to the 1000V dc range

with no input applied. A 20ms ADC measurement is performed and the

result is checked against a limit of 0V

5 mV.

615 Ohms 10 uA source failed This test configures to the 1000V dc

range with the internal 10M 100:1 divider R204 connected across the

input. The 10

A ohms current source is connected. The compliance limit

of the current source is measured. A 20ms ADC measurement is

performed and the result is checked against a limit of 0.12V

0.01V.

616 DC current sense failed This test configures to the 3A dc range. A

20ms ADC measurement is performed and the result is checked against

a limit of 0A

5A. This test confirms that the dc current sense path is

functional. The test limit is set wide because K303

does not open the

current input during self-test. This test should catch a dc current sense

failure without causing false failures when current inputs are applied

during self-test.

617 Ohms 100 uA source failed This test configures to the 1000V dc

range with the internal 10M 100:1 divider R204 connected across the

input. The 100

A ohms current source is connected. The compliance

limit of the current source is measured. A 20ms ADC measurement is

performed and the result is checked against a limit of 0.12V

0.01V.

618 DC high voltage attenuator failed This test configures to the

1000 V dc range. The 500nA ohms current source is connected to

produce a nominal 5V signal. A 20ms ADC measurement is performed

and the result is checked against a limit of 5V

1V.