5 limits test – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 83

84
Figure 4-17
The Remote Interface Operation
You can use the following commands on your PC terminal to make a null
measurement.
CALCulate:FUNCtion NULL
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:NULL:OFFSet {
4.3.5 Limits Test
The limits testing operation allows user to adjust a maximal and a
minimal limit values. The multimeter beeps and an “HI” or “LO” message
will be shown when the reading exceeds the upper or lower limit
respectively. Users can specify the limit values and the values are stored
in a volatile memory. The default values for both upper and lower limits
are “0”. This function is available to all except continuity and diode
measurements.
How to set the limits
You can set the limits or make a limit testing either through the front
panel or the remote interface operation.
Front Panel Operation
To set the limits, user needs to configure the LIMITS math function by
pressing CONFIG + SHIFT + RATIO, and then use ◁ and ▷ to locate
either lower limit (MIN) or upper limit (MAX). Press ENTER to select the