4 other system related operations, 1 display – BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual
Page 90
91
resistance is 600 Ω.
The locations of the buttons are shown with red rectangle frames in
Figure 4-24.
※ Note: Press SHIFT+MIN/MAX again to disable this feature. The
“MATH” anunnciator on the display indicates the state of a mathematical
feature.
Figure 4-24
Remote Interface Operation
Use the following commands to enable dBm feature or to set the
reference resistance:
CALCulate:FUNCtion DBM
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:DBM:REFerence {<value>|MINimum|MAXimum}
4.4 Other System Related Operations
Each system related operation performs a task that is not measurement
related but plays an important role in making your measurements.
4.4.1 Display
1201 has a 5x7 dot matrix, dual-display with three-color (White, Red and
Yellow) annunciators for a better view. A maximum 13 characters are
or