26 stat mode menu (statistical mode), 26 stat mode menu (statistical mode) -47, Boonton 4540 series rf power meter – Boonton 4540 Peak Power Meter User Manual User Manual
Page 95: Operation

Boonton 4540 Series RF Power Meter
Operation
4-47
4.5.26 Stat Mode Menu (Statistical Mode)
The statistical mode menu provides for controlling the horizontal axis scale of the CCDF display, the length of time or
maximum sample count of the CCDF and the action to be taken when the terminal condition is met.
Main Pg1 > Stat Mode > Horiz Scale
Description:
Select the horizontal scale for statistical graphic displays.
Menu Type:
Numeric value.
Range:
0.1 to 5 dB/Div in a 1-2-5 sequence
SCPI Command:
DISPlay:TRACe:HSCALe
Main Pg1 > Stat Mode > Horiz Offset
Description:
Select the horizontal offset for statistical graphic displays.
The value in dBr chosen will appear at the leftmost edge
of the graph.
Menu Type:
Numeric value.
Range:
-50.00 dBr to +50.00 dBr
SCPI Command:
DISPLay:TRACe:HOFFSet
Main Pg1 > Stat Mode > Term Options
Description:
Opens the Terminal Options Menu
Menu Type:
Menu Selection
Main Pg1 > Stat Mode > Cursors
Description:
Opens the statistical Cursors Menu
Menu Type:
Menu Selection
Fig. 4-37 Stat Mode Menu