Status:operation:condition, Status:operation:enable, Status:operation:event – Boonton 4540 Peak Power Meter User Manual User Manual
Page 198: Status:operation:condition? -64, Status:operation:enable -64, Status:operation:event? -64, Boonton 4540 series rf power meter, Remote operation

Boonton 4540 Series RF Power Meter
Remote Operation
5-64
STATus:OPERation:CONDition?
Description:
Return the current value of the Operation Condition register. The following table shows the bit
assignments in the register. These bits are updated by the instrument in real time, and can change
in response to changes in the instrument’s operating condition.
Bit
Value
Definition
0
1
Calibrating
1 = sensor or trigger calibration in progress.
1
2
Not used
always returns 0.
2
4
Not used
always returns 0.
4
16
Measuring
1 = measurement in progress.
5
32
Trigger Status
1 = waiting for a trigger signal.
6
64
Waiting For Arm
1 = waiting for the trigger to be armed.
7
128
Not Used
8
256
Alarm 1
1 = Channel 1 is in an alarm condition.
9
512
Alarm 2
1 = Channel 2 is in an alarm condition.
10
1024
Alarm Latch 1
1 = Channel 1 alarm is latched.
11
2048
Alarm Latch 2
1 = Channel 2 alarm is latched.
12
4096
Not used
always returns 0
13
8192
Not used
always returns 0.
14
16384
Not used
always returns 0.
15
32768
Not used
always returns 0.
Syntax:
STATus:OPERation:CONDition?
Returns:
16-bit register value (0 to 32767)
Valid Modes:
Any
STATus:OPERation:ENABle
Description:
Sets or returns the Operation Enable register, which contains the bit mask that defines which true
conditions in the Operation Status Event register will be reported in the Operation Summary bit of
the instrument Status Byte. If any bit is 1 in the Operation Enable register and its corresponding
Operation Event bit is true, the Operation Status summary bit will be set.
Syntax:
STATus:OPERation:ENABle
Argument:
Valid Modes:
Any
STATus:OPERation:EVENt?
Description:
Returns the current contents of the Operation Event register then resets the register value to 0. The
Operation Event register contains the latched events from the Operation Condition register as
specified by the Operation status group’s positive and negative transition filters.
Syntax:
STATus:OPERation:EVENt?
Returns:
16-bit register value (0 to 32767)
Valid Modes:
Any