Output:external:pulse:perwid, Output:external:pulse:source, Output:external:signal – Boonton 4540 Peak Power Meter User Manual User Manual
Page 174: Output:internal:level[:power, Output:external:pulse:perwid -40, Output:external:pulse:source -40, Output:external:signal -40, Output:internal:level[:power] -40, Boonton 4540 series rf power meter, Remote operation

Boonton 4540 Series RF Power Meter
Remote Operation
5-40
OUTPut:EXTernal:PULse:PERWID
Description:
Set or return the pulse period
control mode of the external 1 GHz calibrator. This parameter has no effect in the preset pulse
control mode or the external pulse source mode.
Syntax:
OUTPut:EXTernal:PULse:PERWID
Argument:
Valid Modes:
Any
OUTPut:EXTernal:PULse:SOURce
Description:
Set or return the pulse modulation source for the external 1 GHz calibrator. If set to INT, the
internal pulse generator is used. EXT selects the signal from the calibrator’s rear-panel EXT
PULSE input. Note this input requires TTL compatible signal levels.
Syntax:
OUTPut:EXTernal:PULse:SOURce
Argument:
Valid Modes:
Any
OUTPut:EXTernal:SIGNal
Description:
Set or return the on/off state of the external 1 GHz calibrator output signal.
Syntax:
OUTPut:EXTernal:SIGNal
Argument:
Valid Modes:
Any
OUTPut:INTernal:LEVel[:POWer]
Description:
Set or return the power level of the internal 50 MHz calibrator output signal.
Syntax:
OUTPut:INTernal:LEVel:POWer < numeric_value >
Argument:
Valid Modes:
Any