Calibration source, External interfaces, Calibration source -9 – Boonton 4540 Peak Power Meter User Manual User Manual
Page 23: External interfaces -9, Boonton 4540 series rf power meter, General information
Boonton 4540 Series RF Power Meter
General Information
1-9
Table 1-2
4540 Series Performance Specifications (continued)
(Specifications are subject to change without notice)
CALIBRATION SOURCE
Internal Calibrator
Operating Modes:
Off, On CW
Frequency:
50.025 MHz ± 0.1%
Level Range:
-60 to +20 dBm
Resolution:
0.1 dB
RF Connector:
Type N
Source VSWR:
1.05 (reflection coefficient = 0.024)
Accuracy, 0C to 20C, NIST traceable:
At 0 dBm: ±0.055 dB (1.27%)
+20 to -39 dBm: ±0.075 dB (1.74%)
-40 to -60 dBm: ±0.105 dB (2.45%)
External Calibrator
Model 2530 1 GHz Calibrator (Optional accessory. See Appendix B)
Auto-calibration:
The Calibrator is used to automatically generate linearity calibration
data for peak power sensors. It is also provides test signals.
EXTERNAL INTERFACES
Remote Control:
Complies with IEEE-488.1 and SCPI version 1993.
GPIB:
Implements AH1, SH1, T6, LE0, SR1, RL1, PP0, DC1, DT1, C0, and E1.
USB:
―USB Device‖ Type B connector - does not require any power from host
LAN:
TCP/IP Ethernet Programmable interfaces;
Multi I/O:
BNC connector, user selectable for status, trigger, alarm or voltage output
Range:
0 to 10V (Analog unipolar), -10V to +10V (Analog bipolar), 0 or 5V (Logic)
Accuracy:
±100mV typical, ±200mV max, uncalibrated
Linearity:
0.1% typical
Ext Cal / VGA Out:
HDB-15 connector. External calibrator control interface for Model 2530, or
video output for VGA compatible analog RGB video monitor.