Read:marker:average, Read:marker:delta, Read:marker:maximum – Boonton 4540 Peak Power Meter User Manual User Manual
Page 183: Read:marker:minimum, Read:marker:average? -49, Read:marker:delta? -49, Read:marker:maximum? -49, Read:marker:minimum? -49, Boonton 4540 series rf power meter, Remote operation

Boonton 4540 Series RF Power Meter
Remote Operation
5-49
READ:MARKer:AVERage?
Description:
For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.
Syntax:
READ[1|2]:MARKer[1|2]:AVERage?
Returns:
CC, average power or voltage at marker
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:MARKer:DELTa?
Description:
For the specified channel return the difference between MK1 and MK2. The units will be the
same as marker units.
Syntax:
READ[1|2]:MARKer:DELTa?
Returns:
CC, (MK1 – MK2)
Where CC is the measurement condition code.
Valid Modes:
All
READ:MARKer:MAXimum?
Description:
For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.
Syntax:
READ[1|2]:MARKer[1|2]:MAXimum?
Returns:
CC, maximum power or voltage at marker
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated.
READ:MARKer:MINimum?
Description:
For the specified channel and marker, return the minimum power or voltage at the marker. The
units will be the same as the specified channel.
Syntax:
READ[1|2]:MARKer[1|2]:MINimum?
Returns:
CC, minimum power or voltage at marker
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated.