FETCh:TEMPerature:AVERage?
Description:
Return the averaged internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.
Syntax:
FETCh[1|2]:TEMPerature:AVERage?
Returns:
CC, Sensor[1|2] Average Temp in degrees C
Where CC is the measurement condition code.
Valid Modes:
Peak Sensors only
FETCh:TEMPerature:CURRent?
Description:
Return the current internal temperature of the specified peak power sensor. Note: If a non-
temperature compensated peak sensor is in use, the 4540 displays a warning message if the
temperature has drifted more than 4C from the autocal temperature.
Syntax:
FETCh[1|2]:TEMPerature:CURRent?
Returns:
CC, Sensor[1|2] Temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
Peak Sensors only
FETCh:TEMPerature:INTernal?
Description:
Return the current internal temperature of the instrument in degrees Celsius.
Syntax:
FETCh:TEMPerature:INTernal?
Returns:
CC, Instrument Temp in degrees C
Where CCn is the measurement condition code.