Boonton 4540 series rf power meter, Remote operation – Boonton 4540 Peak Power Meter User Manual User Manual
Page 219

Boonton 4540 Series RF Power Meter
Remote Operation
5-85
5-1. SCPI COMMAND SUMMARY (continued)
FETCh[1|2]:INTERval:PKAVG?
Return peak to average power between MK1 and MK2.
FETCh:KEY?
Return code of last key depressed
FETCh[1|2]:MARKer[1|2]:AVERage?
Return current reading at the specified marker.
FETCh[1|2]:MARKer:DELTa?
Return difference between MK2 and MK1.
FETCh[1|2]:MARKer[1|2]:MAXimum?
Return maximum reading at the specified marker
FETCh[1|2]:MARKer[1|2]:MINimum?
Return minimum reading at the specified marker
FETCh[1|2]:MARKer:CURsor:PERcent?
Return percent at cursor.
FETCh[1|2]:MARKer:CURsor:POWer?
Return relative power at cursor.
FETCh[1|2]:MARKer:RATio?
Return ratio between MK2 and MK1 as a percentage.
FETCh[1|2]:MARKer:RDELTa?
Return difference between MK1 and MK2.
FETCh[1|2]:MARKer:RRATio?
Return ratio between MK1 and MK2 as a percentage.
FETCh[1|2]:TEMPerature:AVERage?
Return average sensor internal temperature in degrees C
FETCh[1|2]:TEMPerature:CURRent?
Return instantaneous sensor internal temperature in degrees C
FETCh:TEMPerature:INTernal?
Return current internal instrument temperature in degrees C
INITiate:CONTinuous
Set/return state of mode which triggers meas cycles continuously.
INITiate[:IMMediate[:ALL]]
Set mode which starts a measurement cycle when trigger event
occurs
MARKer[1|2]:POSItion:PIXel
Set/return marker position in pixels for selected window.
MARKer[1|2]:POSItion:TIMe
Set/return marker time relative to the trigger.
MARKer:POSItion:PERcent
Set or return the percent probability (y-axis-position) of the CCDF
cursors.
MARKer:POSItion:POWer
Set or return the cumulative relative power (x-axis-position) of the
CCDF cursors.
MEASure[1|2]:POWer?
Return Modulated Mode Pavg in dBm
MEASure[1|2]:VOLTage?
Return Modulated Mode Pavg in equivalent volts.