Read:interval:minfilt, Read:interval:maximum, Read:interval:minimum – Boonton 4540 Peak Power Meter User Manual User Manual
Page 182: Read:interval:pkavg, Read:interval:minfilt? -48, Read:interval:maximum? -48, Read:interval:minimum? -48, Read:interval:pkavg? -48, Boonton 4540 series rf power meter, Remote operation

Boonton 4540 Series RF Power Meter
Remote Operation
5-48
READ:INTERval:MINFilt?
Description:
For the specified channel, return the minimum filtered power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.
Syntax:
READ[1|2]:INTERval:MINFilt?
Returns:
CC, minimum filtered power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:INTERval:MAXimum?
Description:
For the specified channel, return the maximum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax:
READ[1|2]:INTERval:AVERage?
Returns:
CC, maximum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:INTERval:MINimum?
Description:
For the specified channel, return the minimum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax:
READ[1|2]:INTERval:MINimum?
Returns:
CC, minimum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
READ:INTERval:PKAVG?
Description:
For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.
Syntax:
READ[1|2]:INTERval:PKAVG?
Returns:
CC, peak-to-average ratio of power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated