Sense:array:caltemp, Sense:caltemp, Sense:correction:calfactor – Boonton 4540 Peak Power Meter User Manual User Manual
Page 187: Sense:correction:dcycle, Sense:array:caltemp? -53, Sense:caltemp? -53, Sense:correction:calfactor -53, Sense:correction:dcycle -53, Boonton 4540 series rf power meter, Remote operation

Boonton 4540 Series RF Power Meter
Remote Operation
5-53
SENSe:ARRay:CALTemp?
Description:
Returns both channels’ sensor calibration temperatures.
Syntax:
SENSe:ARRay:CALTemp?
Returns:
CC1, Sensor1 cal temp in degrees C, CC2, Sensor2 cal temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
All
SENSe:CALTemp?
Description:
Returns the selected channel’s sensor calibration temperature.
Syntax:
SENSe[1|2]:CALTemp?
Returns:
CC, Sensor[1|2] cal temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
All
SENSe:CORRection:CALFactor
Description:
Set or return the frequency calfactor currently in use on the selected channel. Note setting a
calfactor with this command will override the ―automatic‖ frequency calfactor that was calculated
and applied when the operating frequency was set, and setting the operating frequency will
override this calfactor setting.
Syntax:
SENSe[1|2]:CORRection:CALFactor
Argument:
Valid Modes:
All
SENSe:CORRection:DCYCle
Description:
Set or return the duty cycle correction factor currently in use on the selected channel.
Syntax:
SENSe[1|2]:CORRection:DCYCle
Argument:
Valid Modes:
CW Sensor (Modulated Mode)