Fetch:interval:average, Fetch:interval:maxfilt, Fetch:interval:minfilt – Boonton 4540 Peak Power Meter User Manual User Manual
Page 160: Fetch:interval:maximum, Fetch:interval:average? -26, Fetch:interval:maxfilt? -26, Fetch:interval:minfilt? -26, Fetch:interval:maximum? -26, Boonton 4540 series rf power meter, Remote operation

Boonton 4540 Series RF Power Meter
Remote Operation
5-26
FETCh:INTERval:AVERage?
Description:
For the specified channel, return the average power or voltage in the time interval between marker
1 and marker 2. The units will be the same as the specified channel.
Syntax:
FETCh[1|2]:INTERval:AVERage?
Returns:
CC, average power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
FETCh:INTERval:MAXFilt?
Description:
For the specified channel, return the maximum filtered power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.
Syntax:
FETCh[1|2]:INTERval:MAXFilt?
Returns:
CC, maximum filtered power or voltage between the markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
FETCh:INTERval:MINFilt?
Description:
For the specified channel, return the minimum filtered power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.
Syntax:
FETCh[1|2]:INTERval:MINFilt?
Returns:
CC, minimum filtered power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated
FETCh:INTERval:MAXimum?
Description:
For the specified channel, return the maximum instantaneous power or voltage in the time interval
between marker1 and marker 2. The units will be the same as the specified channel.
Syntax:
FETCh[1|2]:INTERval:MAXimum?
Returns:
CC, maximum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated