Zero Problems
• Perform auto zero, observe the displayed value, it should be stable and lower than +/- 0.5%
• If the result is bad:

Stop flow, set 3.1.1 Low flow cut off to 0, 3.1.3 Flow Mode to “+/-”, perform auto zero,
and totalise over 2 minutes. Compare totalised flow to specified zero stability.


For best process results, zero setting should be performed on process fluid at process temperature.

High Zero’s can be caused by: Leaking valves, Air/Gas inclusions, Coating on tube.

Driver or Sensor Coil Fault
Typical inductance and resistance values:

OPTIMASS 7000

Inductance (mH)

Resistance (Ohm)

OPTIMASS
7000

Driver

Sensor A/B

Driver

Sensor A/B

06/10

5.30 (4.32)

17.32 (10.36)

37 - 42

147 - 152

15

11.7 (8.9)

17.32 (10.36)

47 - 51

147 - 152

25/40

13.1 (11.3)

17.32 (10.36)

40 - 41

147 - 152

50/80

23.5 (12.9)

17.32 (10.36)

49 - 51

147 - 152

• The above data are provided as a rough guide only.
• Damaged magnet coil assembly: Inductance values in brackets.
• Driver = Black and Grey.
• Sensor A = White and Yellow. Sensor B = Green and Purple.
• RTD = Red and Blue (530...550 Ω) at ambient temperature
• Tube strain = Red and Brown:

OPTIMASS 7000 – 06

600 - 800Ω at ambient

OPTIMASS 7000 – 10...80

420 - 560Ω at ambient

• IC strain = Brown and Orange

OPTIMASS 7000 – 06...10

225 - 275 Ω at ambient

OPTIMASS 7000 – 15...80

Not fitted

Resistance values outside these values could indicate a circuit failuire. Meter may be in start-up or
have measuring errors.

All circuit should be isolated from ground (meter case) and each other: >20MΩ. If circuits are shorting
to ground, meter may be in start-up.

Caution:
Fluid may be in secondary containment – possible tube failure. Depressurise and safely remove from line
as soon as possible.

OPTIMASS 3000 (7100)

Inductance (mH)

Resistance (Ohm)

OPTIMASS
3000 (7100)

Driver

Sensor A/B

Driver

Sensor A/B

01

1.2 (1.2)

7.2 (7.2)*

54 – 60

105 - 110

03/04

2.6 (8.9)

10.5 (10.36)

43 – 50

132 - 138


• The above data are provided as a rough guide only.
• Damaged magnet coil assembly: Inductance values in brackets.
• Driver = Purple/Black and Orange/Grey.
• Sensor A = White and Yellow. Sensor B = Green and Yellow.
• RTD = Red and Blue (530...550 Ω) at ambient temperature
• Resistance values outside these values could indicate a circuit failure. Meter may be in start-up or have

measuring errors.

• All circuit should be isolated from ground (meter case) and each other: >20MΩ. If circuits are shorting

to ground, meter may be in start-up.

Caution:
Fluid may be in secondary containment – possible tube failure. Depressurise and safely remove from line
as soon as possible.