Zero Problems
• Perform auto zero, observe the displayed value, it should be stable and lower than +/- 0.5%
• If the result is bad:
Stop flow, set 3.1.1 Low flow cut off to 0, 3.1.3 Flow Mode to “+/-”, perform auto zero,
and totalise over 2 minutes. Compare totalised flow to specified zero stability.
For best process results, zero setting should be performed on process fluid at process temperature.
High Zero’s can be caused by: Leaking valves, Air/Gas inclusions, Coating on tube.
Driver or Sensor Coil Fault
Typical inductance and resistance values:
OPTIMASS 7000
Inductance (mH)
Resistance (Ohm)
OPTIMASS
7000
Driver
Sensor A/B
Driver
Sensor A/B
06/10
5.30 (4.32)
17.32 (10.36)
37 - 42
147 - 152
15
11.7 (8.9)
17.32 (10.36)
47 - 51
147 - 152
25/40
13.1 (11.3)
17.32 (10.36)
40 - 41
147 - 152
50/80
23.5 (12.9)
17.32 (10.36)
49 - 51
147 - 152
• The above data are provided as a rough guide only.
• Damaged magnet coil assembly: Inductance values in brackets.
• Driver = Black and Grey.
• Sensor A = White and Yellow. Sensor B = Green and Purple.
• RTD = Red and Blue (530...550 Ω) at ambient temperature
• Tube strain = Red and Brown:
OPTIMASS 7000 – 06
600 - 800Ω at ambient
OPTIMASS 7000 – 10...80
420 - 560Ω at ambient
• IC strain = Brown and Orange
OPTIMASS 7000 – 06...10
225 - 275 Ω at ambient
OPTIMASS 7000 – 15...80
Not fitted
•
Resistance values outside these values could indicate a circuit failuire. Meter may be in start-up or
have measuring errors.
•
All circuit should be isolated from ground (meter case) and each other: >20MΩ. If circuits are shorting
to ground, meter may be in start-up.
Caution:
Fluid may be in secondary containment – possible tube failure. Depressurise and safely remove from line
as soon as possible.
OPTIMASS 3000 (7100)
Inductance (mH)
Resistance (Ohm)
OPTIMASS
3000 (7100)
Driver
Sensor A/B
Driver
Sensor A/B
01
1.2 (1.2)
7.2 (7.2)*
54 – 60
105 - 110
03/04
2.6 (8.9)
10.5 (10.36)
43 – 50
132 - 138
• The above data are provided as a rough guide only.
• Damaged magnet coil assembly: Inductance values in brackets.
• Driver = Purple/Black and Orange/Grey.
• Sensor A = White and Yellow. Sensor B = Green and Yellow.
• RTD = Red and Blue (530...550 Ω) at ambient temperature
• Resistance values outside these values could indicate a circuit failure. Meter may be in start-up or have
measuring errors.
• All circuit should be isolated from ground (meter case) and each other: >20MΩ. If circuits are shorting
to ground, meter may be in start-up.
Caution:
Fluid may be in secondary containment – possible tube failure. Depressurise and safely remove from line
as soon as possible.