Installation and Operating Instructions OPTIMASS

14

1.2.7 Heating and insulation

There are several methods to heat the meter. In most cases heating is unnecessary as the meter is

designed as such that very little heat is lost or gained through the outer cylinder.


Insulation

Where insulation is required a variety of materials may be used to insulate the meter. Care must be

taken not to insulate the meter above the halfway mark of the electronics support post as shown in the

sketch.

Do not insulate above

this line.

Insulation


Electrical Heating

Electrical tape heating may be used. Care should be taken to only heat the sections where the best

effect will be achieved. Do not heat above the converter mount centre line as shown above.

The following guidelines must be observed.





Areas A and B may be

heated. Area C must not be

heated.

When insulating please observe guidelines as per insulation section.


DIM A and B

Size

Titanium

Hastelloy + SS 318

10 50 -

15 65 65

25 120 75

40 150 150

50 200 125

80 410 225