6.2 Error Messages
• Basic Errors: These errors are displayed independently of which error message function is selected.
• Transducer Errors: These errors only occur when error massage function is set to TRANS.ERROR or
ALL ERRORS
• I/O Errors: These errors only occur when TRANS.ERROR or ALL ERRORS is set to I/O Errors or All
Errors
• All Errors: All errors can occur
• Errors are encoded as follows:
Bit
Error Name
Error definition
Error type
Error Level
0
MASS FLOW
Measured mass flow value over range
Basic error
Light
1
ZERO ERROR
Excessive flow measured during zero calibration
Basic error
Light
2
TOTAL O/F
Fixed precision totaliser has rolled over
Basic error
Light
3
Not used
4
Temperature
Temperature is outside the operating ranges
Basic error
Light
5
Sensor A
Sensor A voltage signal is less than 5% of desired value
Transducer error
Light
6
Sensor B
Sensor B voltage signal is less than 5% of desired value
Transducer error
Light
7
Ratio A/B
One sensor signal is much larger than the other
Basic error
Severe
8
DC A
DC voltage part of sensor A is larger than 20% ADC
Basic error
Severe
9
DC B
DC voltage part of sensor B is larger than 20% ADC
Basic error
Severe
10
Not used
11
Sampling
No synchronization with primary head
Basic error
Severe
12
Not used
13
ROM DEFAULT
EEPROM checksum error detected on start up.
Default data has been loaded.
Basic error
Severe
14
Not used
15
EEPROM
Unable to save data to EEPROM. Hardware fault
Basic error
Fatal
16
NVRAM
Checksum error detected on startup. Previous data lost
Basic error
Severe
17
NVRAM FULL
NVRAM has exceeded 50,000 Cycles
Basic error
Fatal
18
POWER.FAIL
Custody transfer only. There has been an interruption of
power to the converter.
Basic error
Light
19
Watchdog
Converter reset by the watchdog. Last NVRAM save failed.
Basic error
Fatal
20
Not used
21
Temp Custody
Temperature has drifted by 30 degrees from the zero
calibration temp
Basic error
Light
22
RESIST.CIR
Resistance circuit has failed
Basic error
Light
23
I 1 SAT.
Current output 1 is outside the set ranges
I/O Error
Light
24
FREQ SAT.
Frequency/pulse output is outside the set ranges
I/O Error
Light
25
ALARM.OUT.A
Alarm output is outside the set ranges
I/O Error
Light
26
I 2 SAT
Current output 2 is outside the set ranges
I/O Error
Light
27
I 3 SAT
Current output 3 is outside the set ranges
I/O Error
Light
28
COMM.FAIL
Communication failure > 5 attempts without valid response
Basic error
Severe
29
SYS Changed
The front end or backend do not match. (One has been
changed
Basic error
Light
30
SYSTEM
Converter reset
Basic error
Severe
31
Not used
Errors on POWER UP
•
WIRING ERROR : Incorrect wiring detected on remote meters
•
FRONT END ERROR :