Installation and Operating Instructions OPTIMASS

94

6.2 Error Messages

• Basic Errors: These errors are displayed independently of which error message function is selected.
• Transducer Errors: These errors only occur when error massage function is set to TRANS.ERROR or

ALL ERRORS

• I/O Errors: These errors only occur when TRANS.ERROR or ALL ERRORS is set to I/O Errors or All

Errors

• All Errors: All errors can occur
• Errors are encoded as follows:

Bit

Error Name

Error definition

Error type

Error Level

0

MASS FLOW

Measured mass flow value over range

Basic error

Light

1

ZERO ERROR

Excessive flow measured during zero calibration

Basic error

Light

2

TOTAL O/F

Fixed precision totaliser has rolled over

Basic error

Light

3

Not used

4

Temperature

Temperature is outside the operating ranges

Basic error

Light

5

Sensor A

Sensor A voltage signal is less than 5% of desired value

Transducer error

Light

6

Sensor B

Sensor B voltage signal is less than 5% of desired value

Transducer error

Light

7

Ratio A/B

One sensor signal is much larger than the other

Basic error

Severe

8

DC A

DC voltage part of sensor A is larger than 20% ADC

Basic error

Severe

9

DC B

DC voltage part of sensor B is larger than 20% ADC

Basic error

Severe

10

Not used

11

Sampling

No synchronization with primary head

Basic error

Severe

12

Not used

13

ROM DEFAULT

EEPROM checksum error detected on start up.
Default data has been loaded.

Basic error

Severe

14

Not used

15

EEPROM

Unable to save data to EEPROM. Hardware fault

Basic error

Fatal

16

NVRAM

Checksum error detected on startup. Previous data lost

Basic error

Severe

17

NVRAM FULL

NVRAM has exceeded 50,000 Cycles

Basic error

Fatal

18

POWER.FAIL

Custody transfer only. There has been an interruption of
power to the converter.

Basic error

Light

19

Watchdog

Converter reset by the watchdog. Last NVRAM save failed.

Basic error

Fatal

20

Not used

21

Temp Custody

Temperature has drifted by 30 degrees from the zero
calibration temp

Basic error

Light

22

RESIST.CIR

Resistance circuit has failed

Basic error

Light

23

I 1 SAT.

Current output 1 is outside the set ranges

I/O Error

Light

24

FREQ SAT.

Frequency/pulse output is outside the set ranges

I/O Error

Light

25

ALARM.OUT.A

Alarm output is outside the set ranges

I/O Error

Light

26

I 2 SAT

Current output 2 is outside the set ranges

I/O Error

Light

27

I 3 SAT

Current output 3 is outside the set ranges

I/O Error

Light

28

COMM.FAIL

Communication failure > 5 attempts without valid response

Basic error

Severe

29

SYS Changed

The front end or backend do not match. (One has been
changed

Basic error

Light

30

SYSTEM

Converter reset

Basic error

Severe

31

Not used


Errors on POWER UP

WIRING ERROR : Incorrect wiring detected on remote meters

FRONT END ERROR :