KROHNE OPTIMASS with MFC 05x EN User Manual
Page 66
Installation and Operating Instructions OPTIMASS
65
Fct.
No
Text Description
and
settings
4.8
SYS.CTROL
Submenu 4.8 System control
4.8.1 FUNCTION
Function for system control
* OFF (System control inactive)
* FLOW = 0 (Mass flow readings forced to zero, totaliser frozen)
* FLOW = 0/RST. (Mass flow readings forced to zero, totaliser frozen while active
but reset to zero as condition becomes inactive. Not available with Custody
Transfer Protection)
* OUTPUTS.OFF (Forces all outputs to their OFF states)
4.8.2 CONDITION
Condition for triggering the above function
* DENSITY (Function is triggered if density goes outside Max or Min limits as set in
Fcts 4.8.3 and 4.8.4)
* TEMPERATUR (Function is triggered if temperature goes outside Max or Min
limits as set in Fct 4.8.3 and 4.8.4).
Function not available with Custody Transfer Protection.
4.8.3 LOW
LIMIT
Minimum allowable value of temperature or density selected in Fct. 4.8.2
Units: depend on function but will correspond to those set in Fct. 3.2.6 and 3.2.7
Function not available with Custody Transfer Protection.
4.8.4 HIGH
LIMIT
Maximum allowable value of temperature or density selected in Fct.4.8.2
Units: depend on function but will correspond to those set in Fct. 3.2.6 and 3.2.7
Function not available with Custody Transfer Protection
4.9
COMM.MODULE
Submenu 4.9 Communication modules
4.9.1
PROTOCOL
Display of communication protocol fitted
(OFF, SERIAL, HART, MODBUS, PROFIBUS, FF BUS or KROHNE)
4.9.2 ADDRESS
Address
(not available for selection OFF and SERIAL in Fct. 4.9.1)
4.9.3
BAUDRATE
Setting Baudrate (for option MODBUS in Fct. 4.9.1 only)
4.9.4
SER.FORMAT
Serial format (for option MODBUS in Fct. 4.9.1 only)
4.10
CALIB I
Submenu 4.10 Calibration current output 1
4.10.1 I 1 5 mA
Calibration of the current output 1 for 5 mA
4.10.2 I 1 18 mA
Calibration of the current output 1 for 18 mA
4.10.3 I 2 5 mA
see I 1 5 mA
4.10.4 I 2 18 mA
see I 1 18 mA
4.10.5 I 3 5 mA
see I 1 5 mA
4.10.6 I 3 18 mA
see I 1 18 mA